8.0.2 The Fallacy of the "Average Person" 8.1.2 Nonspecific Analysis This problem of nonspecificity is most noticeable in the use of
infrared breath analyzing instruments, such as the various Intoxilyzer models,
the Intoximeter 3000 and the BAC Datamaster. CMI's so-called
"state-of-the-art" Intoxilyzer Model 5000 (and its various
permutations), the most widely used breath machine in service today, utilizes
infrared spectroscopy. Yet infrared analysis is particularly susceptible to
giving false readings due to nonspecificity. In fact, the single greatest flaw
in the Intoxilyzer itself—and the most productive area for cross-examination—is
the machine's inherent lack of specificity. The technical reason for this lack
of specificity is that the Intoxilyzer is not designed to detect the molecule
of ethyl alcohol (ethanol), but rather only a part of that molecule—the methyl
group. In other words, it is the methyl group in the ethyl alcohol compound
that is absorbing the infrared light, resulting in the eventual blood-alcohol
reading. Thus the machine will "detect" any chemical compound and
identify it as ethyl alcohol if it contains a methyl group compound within its
molecular structure. The Intoxilyzer assumes that the methyl group is a part of
an ethyl alcohol compound. The simple fact is that there are numerous compounds that contain
the methyl group. Among these are isopropyl alcohol, propane, butane,
propylene, methane, ethane, ethyl chloride, acetic acid, butadiene,
dimethylether, dimethylamine and dimethylhydrazine. Acetone and acetaldehyde
both contain the methyl group in their structure—and, interestingly, each can
be found on the human breath. In fact, recent studies have found that over one
hundred chemical compounds can be found on the breath at any given moment in
time (see Table 1). More important, approximately 70 to 80 percent of these
compounds contain methyl groups. And the infrared breath machine will detect
each of these as "ethyl alcohol." To make matters worse, the machine detects alcohol through
"additive absorption." In other words, the more methyl groups the instrument
detects by their absorbing the infrared energy, the higher will be the
blood-alcohol reading. Thus all of the non-alcoholic compounds on the breath
will have a cumulative effect—that is, the errors will be added one on top of
another. To approach this problem of specificity from another angle,
ethanol has a peak absorption at approximately 3.39 microns, with the band of
absorption declining rapidly on each side. However, ethanol has wide absorption
bands and peaks at other wavelengths as well, such as bands in the 7.25, 9.5
and 11.0 micron range. Yet the Intoxilyzer screens out these wavelengths and
attempts to identify the substance, ethanol, on the basis of only one, two, or
three bands, depending on the model. This reduces the efficiency of the Intoxilyzer,
since other elements besides ethanol absorb in these ranges, but few, if any,
of them would also absorb in the 7.25, 9.5 and 11.0 micron ranges. To make this point conceptually simpler, the analogy of
fingerprint identification has been used. Ethanol has a "fingerprint"
in that it will absorb energy wavelengths of roughly 3.39, 7.25, 9.5 and 11.0
microns. While many other substances have fingerprints that will include the
3.39 range or the 7.25 range or the 11.0 range, probably none of them will have
all of them—that is, the full print of ethanol. By fingerprint analogy, then,
absorption of light wavelengths at about 3.39 microns represents a single
"point of identity"; absorption at several bands would represent more
points of identity. Certainly, fingerprint identification is considered more positive
because there are many points of identity and, conversely, availability of only
one, two, or three points of identity in finger. Table 1 1. Acetone 52. 5-Ethyl-l-butanol Printing would preclude an expert's opinion as to similarity. Yet
the Intoxilyzer attempts to identify a substance on the basis of one, two, or
three points of identity, despite the availability of other points that would
exclude all substances other than ethyl alcohol. How prevalent are chemicals in the breath that can register on
breath analyzing machines as ethanol? There have been a number of recognized
studies on the existence of chemical compounds on the breath — all concluding
that a wide variety of compounds exists, including compounds containing the
methyl group. The results of one such study are found in an article by Conkle,
Camp and Welch entitled Trace Composition of Human Respiratory Gas, 30 Archives
of Environmental Health 290 (1975). The researchers analyzed the breath of
eight test subjects and found "the presence of 69 different compounds in
the expired air of eight men." Id. at 292. Another article, by
Krotoszynski, Gabriel and O'Neill, Characterization of Human Expired Air: A
Promising Investigative and Diagnostic Technique, 15 Journal of Chromatographic
Science 240 (1977), described analysis of air samples taken from 28
"average" human subjects. This study found that the "combined
expired air comprises at least 102 various organic compounds of endogenous and
exogenous origin." Id. at 244. The researchers further concluded that
"400% of the constituents (70% of the mean organic contents) are common to
76% of the population studied." Id. at 244. Finally, Canadian scientists
have reported that "approximately 200 compounds have been detected in the
human breath." Manolis, The Diagnostic Potential of Breath Analysis, 29(1)
Clinical Chemistry 5 (1983). This last study confirmed the presence of acetone on the breath in
diabetics and in persons on a diet "associated with a weight reduction of
about one-half pound per week." Id. at 9. Another study has confirmed that
diabetics may give false indications of intoxication. In Brick, Diabetes, Breath
Acetone and Breathalyzer Accuracy: A Case Study, 9(1) Alcohol, Drugs and
Driving (1993), a researcher found that expired ketones in the breath of an
untreated diabetic can contribute to erroneously high breath-alcohol readings.
Further, the acetone on the breath from ketoacidosis will result in an odor of
alcohol. Finally, behavioral patterns of a diabetic whose blood-sugar level has
dropped will include slurred speech, slow gait, impaired motor control,
fumbling hand movements and mental confusion—all symptomatic of intoxication. Acetone may also be found on the breath of perfectly normal,
healthy individuals. Yet, acetone is one of the compounds that will be detected
on many breath analyzing instruments as ethanol. In the Intoxilyzer, for
example, it is detected because acetone absorbs infrared energy in the 3.38 to
3.40 micron range—the same range where ethanol is found. Therefore, if acetone
were introduced into the Intoxilyzer, the machine would simply register the
presence of alcohol despite its absence. If an individual had 525 micrograms
per liter of acetone in the breath, he would register a blood-alcohol level of
.02 to .03 percent. Thus, if an individual with a true blood-alcohol level of
.08 percent had that amount of acetone, the Intoxilyzer would register in the
area of .l0 to .11 percent. The National Highway Traffic Safety Administration
has published a report entitled The Likelihood of Acetone Interference in
Breath Alcohol Measurement (DOT HS—806-922). The report basically summarizes
scientific literature on the subject, concluding that normal individuals have
insignificant levels of acetone on their breath. The data indicated, however,
that dieters can have higher levels and that diabetics not in control of their
blood-sugar had levels hundreds or even thousands of times higher than normal.
Unfortunately, the authors did not determine what effect such levels would have
on a breath testing device; they simply concluded that at levels rendering the
individual "not too ill to drive," the breath reading would be raised
by only .01-.02 percent. The authors (and it must be recognized that this federal
agency has been consistently law enforcement-oriented, to the point of
suppressing unfavorable results of radio frequency interference tests, for
example) also conclude that the only instrument significantly affected by
acetone interference is the Intoxilyzer 4011A. By contrast, a more scientific article entitled Excretion of
Low-Molecular Weight Volatile Substances in Human Breath: Focus on Endogenous
Ethanol, 9 Journal of Analytical Toxicology 246 (1985), has concluded that
acetone can exist in some normal individuals in quantities that can create
falsely high results in a breath-alcohol test. For a study confirming the
effects of increased levels of acetone in dieters, see Frank and Flores, The
Likelihood of Acetone Interference in Breath Alcohol Measurement, 3 Alcohol,
Drugs and Driving 1 (1987). In that study, researchers found that fasting can
increase acetone to levels sufficient to obtain readings of .06 percent on
breath testing instruments. Similarly, a study confirming the effects of acetone in diabetics
can be found in Mormann, Olsen, Sakshaug and Morland, Measurement of Ethanol by
Alkomat Breath Analyzer; Chemical Specificity and the Influence of Lung
Function, Breath Technique and Environmental Temperature, 25 Blutalkohol 153
(1988). Diabetic subjects in that study also were found to have acetone levels
sufficient to produce breath-alcohol readings of .06 percent. Another major source of nonspecific detection in various methods
of blood-alcohol analysis involves the presence in the body of acetaldehyde.
Acetaldehyde is a chemical by-product in the body's metabolism of ethanol. As
the ethanol is oxidized, or "burned off," acetaldehyde is produced;
eventually, the acetaldehyde is converted to carbon dioxide and water. This oxidation of alcohol with its attendant production of
acetaldehyde takes place primarily in the liver. However, it can also occur in
other tissues and fluids of the body—most notably in the lungs. And, since
acetaldehyde, like acetone, will be "detected" as ethanol in many
types of analysis, a falsely high blood-alcohol reading can result. This
nonspecific detection of acetaldehyde in the breath is particularly noticeable
in infrared spectroscopy, "wet chemical" analysis and blood analysis
using the oxidation technique. In the past, this production of acetaldehyde has not been seen to
be a problem. When acetaldehyde is produced in the liver, it is sent into the
blood and then into the lungs in small quantities. More recently, however, it
has been discovered that alcohol is metabolized—and acetaldehyde produced—in
the lungs themselves. The result can be a highly elevated amount of
acetaldehyde in the lungs, with a subsequently large amount in the expired
breath to register in a breath analyzing instrument as ethanol. This phenomenon of production of acetaldehyde in the lungs was
commented on in a study by Jauhanen, Baraona, Hiyakawa and Lieber, entitled
Origin of Breath Acetaldehyde During Ethanol Oxidation: Effect of Long-Term
Cigarette Smoking, 100 Journal of Laboratory Clinical Medicine 908 (1982). The
researchers discovered that the amount of acetaldehyde in the lungs was
considerably greater than the amount that would be expected if passed from the
liver by way of the blood into the lungs. Furthermore, the elevated amounts of acetaldehyde
in the lungs were not predictable—they varied according to the individual.
Thus, for example, it was found that acetaldehyde concentrations in the lungs
were far greater for smokers than for nonsmokers. Translated into practical
effect, smokers are more likely to have high blood-alcohol readings, regardless
of their true blood-alcohol level. In a subsequent study, it was found that breath acetaldehyde
levels were found to indicate blood-alcohol levels 30 times higher than would
be expected from direct blood analysis. Stowell, et al., A Reinvestigation of
the Usefulness of Breath Analysis in the Determination of Blood Acetaldehyde
Concentration, 8(5) Alcoholism: Clinical and Experimental Research 442 (1984).
The conclusion of the researchers was, again, that the acetaldehyde in the
lungs was not coming from the liver by way of the blood, but was being produced
in the lungs themselves and exhaled in much larger quantities than would be
expected. End result: falsely high breath test readings. Another example of variances in acetaldehyde levels that can
compound blood-alcohol analysis was reported by researchers studying
alcoholics. In Lindros, et al., Elevated Blood Acetaldehyde in Alcoholics and
Accelerated Ethanol Elimination, 13 (Supp. 1) Pharmacology, Biochemistry and
Behavior 119 (1980), scientists discovered that acetaldehyde in the breath and
blood of alcoholics was 5 to 55 times higher than that in nonalcoholics. Thus
increased acetaldehyde—and consequent falsely high blood-alcohol readings—can be
attributed to the makeup of the alcoholic's physiology. Of course, counsel
should consider the risks in bringing this information out for the jury to
ponder. Thus it clearly appears that acetaldehyde is produced in the lungs
themselves. With this elevated level of acetaldehyde in the lungs, there will
be elevated levels in the breath. And apparently every breath analyzing device
but those employing gas chromatography is fully capable of registering this
compound as ethanol. Further, blood tests utilizing the process of oxidation
reaction with potassium dichromate can be affected by acetaldehyde. In their
latest advertisements for the Intoxilyzer Model 5000, the manufacturer lists an
acetaldehyde detector as a new option. In developing this option, the manufacturer
has clearly acknowledged the often-denied problem of acetaldehyde interference.
Whether the device is effective is, of course, another matter. Certainly,
counsel confronted with a case involving the Model 5000 should determine
whether the option was present; if not, that fact should be pointed out to the
jury. In fact, the existence of such a device in this
"state-of-the-art" instrument should be interesting to a jury in DUI
cases involving any type of breath testing apparatus. It is also interesting to note that the commonly observed symptoms
of flushed face and bloodshot eyes are physiologically the result of the
effects of acetaldehyde—not of alcohol itself. Thus such symptoms may actually
tend to corroborate the falsity of a blood-alcohol test—that is, elevated
acetaldehyde evidenced by a flushed face and bloodshot eyes may have caused the
high breath test reading! An interesting study involving the Intoxilyzer 4011-AS has shown
dramatically the dangers of nonspecificity. In an article appearing in 7(4)
Drinking/Driving Newsletter 3 (February 19, 1988), a test conducted by the
Demers Laboratory in Springvale, Maine, is described wherein a subject was
tested after exposure under realistic field conditions to paint and glue. The
subject entered a test room and applied a pint of contact cement to a piece of
plywood; he then applied a gallon of oil-base paint to a vertical surface. This
activity lasted about one hour. Twenty minutes after leaving the room, the subject was tested on
the Intoxilyzer. Results? Despite the subject's having no alcohol in his body,
the machine registered .12 percent—over the legal limit. The subject was tested
again one-half hour later: Readings of .05 and .04 percent were obtained. Another example of commonly encountered chemical compounds that
can affect breath tests was described in a study reported in Giguiere, Lewis,
Baselt and Chang, Lacquer Fumes and the Intoxilyzer, 12 Journal of Analytical
Toxicology 168 (1988). Scientists performed tests on a professional painter who
was exposed to lacquer fumes under controlled conditions. In the first test, he
sprayed paint in a room for 20 minutes, wearing a protective mask; his blood
and breath were then tested. Although the blood test showed no presence of
alcohol, an Intoxilyzer 4011-AS indicated a reading of .075 percent BAC. Ten minutes later, the painter sprayed the same room for five
minutes—but this time without the protective mask. The blood test again showed
no BAC. The Intoxilyzer, however, registered a reading of .48 percent! Perhaps
most interesting, the Intoxilyzer was equipped with an acetone detection light
designed to detect the presence of any interfering compounds — yet at no time
during the test did the light indicate the presence of any such compounds. In an article appearing on the front page of the August 24, 1988,
edition of the Spokane Spokesman Review, an individual in a Sandpoint Idaho
jail awaiting trial for drunk driving claimed that he had been siphoning
gasoline; when he sucked on the hose, he swallowed some of the gasoline, which
later caused a high reading on a breath test. He managed to talk the sheriff
into a demonstration to prove his story. Taken from his cell after one week of
incarceration, he swallowed a cup of unleaded gasoline; after various periods
of time, he blew into an Intoximeter. The results? After 5 minutes, the reading was .00 percent; after
10 minutes, .04 percent; after 20 minutes, the machine registered .31 percent;
and after one hour, the reading was .28 percent. Three hours after ingestion,
the individual still blew a .24 percent on the Intoximeter! A quick call to a
gasoline distributor confirmed that gasoline contains no alcohol. This phenomenon has been scientifically verified in a study
conducted by CMI, the manufacturer of the Intoxilyzer and reported in 8(3)
Drinking/Driving Law Letter 6 (1989). The CMI technicians mixed a simulator
solution of 800 micrograms of gasoline with 500 milliliters of distilled water,
then introduced it into an Intoxilyzer 5000. The solution produced readings of
.619 percent, .631 percent and .635 percent. Diethyl ether is yet another substance that will be falsely
detected as "alcohol" by breath testing instruments. The compound,
which can be inhaled, is found in some automotive products and is used in the
manufacture of plastics and smokeless gunpowder. See Bell, et al., Diethyl
Ether Interference with Infrared Breath Analysis, 16 Journal of Analytical
Toxicology (1992), for a study that concluded that ''[t]he possibility of
interference with an alcohol reading by ether or by other substances may
therefore render prosecution more difficult, if not impossible." In another study, researchers discovered that three other
compounds found in common products falsely registered as ethyl alcohol on the
machines. Cowan, et al., The Response of the Intoxilyzer 4011ASA to a Number of
Possible Interfering Substances, 35(4) Journal of Forensic Sciences 797 (1990).
One of the substances, methyl ethyl ketone, is used in lacquers, paint
removers, cements and adhesives, celluloid and cleaning fluids. Another,
toluene, is used in paints, lacquers, varnishes and glues. The third is
isopropanol, commonly known as rubbing alcohol. Phil Price, a nationally prominent DUI attorney in Montgomery,
Alabama, conducted a series of experiments in which subjects ingested various
foods and were then tested on an Intoxilyzer 5000 (64-series). Interestingly,
bread caused the highest readings! Using alcohol-free subjects, Price
consistently obtained readings in the area of .05 percent after consumption of
various types of bread products. Further, the slope detector failed to detect
any interferent during the tests.
One of the greatest sources of error in blood-alcohol testing is the consistently recurring fallacy that the individual tested is perfectly average in certain critical physiological traits. Put another way, obtaining an accurate blood-alcohol reading is completely dependent on the validity of a number of scientific assumptions. Unfortunately for the person being tested, these assumptions are usually incorrect: The person tested is rarely "average" in even one of these critical characteristics, let alone in all of them.
Counsel in a DUI case will constantly be confronted by these almost hidden assumptions. And it is very important that these false premises be brought out for the jury—along with the fact that the final readings fall with the presumptions.
Thus, for example, all breath testing devices depend on the assumption that the ratio between alcohol in the exhaled breath and alcohol in the blood is 1 to 2100. In fact, the machine is designed to produce a reading based on that assumption; the accuracy of the reading is directly tied to the accuracy of the presumption. Yet, as will be discussed more fully in 8.1.1 the actual ratio in any given individual can vary from 1:1300 to 1:3000, or even more widely. Thus a person with a true blood-alcohol level of .08 but a breath-to-blood ratio of 1:1700 would have a .10 reading on an "accurate" breath testing instrument.
Put simply, these machines do not test individuals. Rather, they test the average person over and over again, but using the subject's breath.
Yet another example of the assumption of "averageness" can be found in urinalysis. When a subject's urine is analyzed for blood-alcohol, a presumption exists that there are 1.3 parts of alcohol in the bladder's urine for every 1 part of alcohol in the blood. This 1:1.3 ratio is as fallacious as the 1:2100 ratio—that is, it is based entirely on the ratio found in the average person. As is discussed more fully in 8.4.1 however, the actual ratio found in any given individual can vary greatly. And as the ratio is in error, so will be the final blood-alcohol reading.
Another example of this constant reliance on averages shows itself when the prosecutor offers evidence of retrograde extrapolation (see 8.0.6) The blood-alcohol level at the time of testing is not relevant to the charge, of course and so the state will offer evidence to show what the level was when the defendant was driving. This is commonly done by extrapolating backward—that is, computing the earlier blood-alcohol level by estimating how much alcohol had been eliminated or "burned off" in the interim between driving and testing. But this requires two assumptions: The blood-alcohol level was declining and the rate of elimination is known. This second assumption involves the further assumption that the "burn-off" rate was .015 percent per hour. How does the prosecution know that the defendant was eliminating (assuming he was eliminating) at that rate and not at .005 percent or .3 percent! Quite simply, the prosecution does not know: It merely assumes that the defendant eliminates at the average rate. And, of course, error in such an assumption translates into error in the extrapolation.
This ubiquitous "average person" in the DUI arena is not limited to chemical analysis. We even find him with the arresting officer in the field. When the officer administers the increasingly common "horizontal gaze nystagmus'' test as part of the battery of field sobriety tests, he operates on the assumption that the suspect is "Mr. Average." As has been discussed in 7.4, the officer has been trained to "read" at what angle the suspect's eyes begin jerking. A blood-alcohol reading can theoretically be obtained by subtracting the angle from 50; jerking at 35 degrees, for example, would mean the suspect has a blood-alcohol level of .15 percent. Where does the magic figure of 50 come from? The average person.
An alternative method of administering the nystagmus test is to "flunk" the person if jerking begins before 40 or 45 degrees. Why? Again, because the average person would theoretically have .10 or .05 percent alcohol in his blood at this point.
In either test, of course, we do not know what the individual's actual "baseline" is—that is, the angle at which his eyes would begin jerking if he were sober. In both cases, the individual is assumed to be physiologically identical to the theoretical "average" person.
One of the major defects in many methods of blood-alcohol
analysis is the failure to identify ethanol (also referred to as ethyl alcohol)
to the exclusion of all other chemical compounds. To use the terminology of
scientists, such methods are not specific for ethanol: They will detect other
compounds as well, identifying any of them as "ethanol." Thus a
client with other compounds in his blood or breath may have a high
"blood-alcohol" reading with little or no ethanol in his body.
Normal Composite Compositional Profile of Human Expired Air
2. Isoprene
3. Acetonitrile
4. pTolualdehyde
5. Toluene
6. P,SDimethylhexane
7. Ethyl Alcohol
8. Acetaldehyde
9. Dichloronitromethane
10. 2,2,4-Trimethyl-l-pentanol
11. n-Propyl acetate
12. 2,2-Dimethyl-l-pentanol
13. Cyclohexane
14. Hexane
15. Thiolacetic acid
16. I-Heptanol
17. Cyclohexyl alcohol
18. Benzene
19. 2-Ethyl-l-hexanone
20. 2,3,5 Trimethylhexane
21. Ethyl Imercaptopropionate
22. Cycloheptatriene
23. p-Xylene
24. n-Butyl alcohol
25. 3,4 Dimethylhexane
26. Limonene
27. Isooctyl alcohol
28. Methyl-n-propyl sulfide.
29. 2 - Ethyl-4-methyl-1-pentanol
30. Neopentyl acetate
31. Trans4nonenal
32. n-Heptane
33. Ethylbenzene
34. 5-Methyl4heptanone
35. Dimethylsulfide
36. P-Methyl-l-pentanol
37. pl)ichlorobenzene
38. Trans-3-hexen-l-ol
39. Capryl alcohol
40. Mesitylene
41. n-Hexylmercaptan
42. 3,4-Dimethylheptane
43. 2,3,3,4-Tetramethylpentane
44. 1Chlorohexane
45. Dichloroacetylene
46. 2,P-Dimethyl-l-octanol
47. 2,2,3,3 - Tetramethylhexane
48. o-Xylene
49. 2,3,3 - Trimethylhexane
50. Isopropylalcohol
51. 2,2-Dimethyl-l-hexanol
53. Z,P-Dimethylheptane
54. Furan
55. Naphthalene
56. Thiocyclopentane
57. Cyclopentylalcohol
58. n-l\lonane
59. Ethyl phenyl acetate
60. n-Amyl alcohol
61. Z,CDimethylheptane
62. 5-Nitropropane
63. 2,6 - Di-tert-butyl-4-methyl-phenol
64. Methyl-tert-butyl-ketone
65. Di-Tert-butyldisulfide
66. 2,2-Dimethyl-Shexanone
67. 1,2-Diethylbenzene
68. 2,5-Dimethylheptane
69. 2-Methyl-3-heptanone
70. Isobutyl alcohol
71. m-Xylene
72. 2,2,5,5Tetramethylhexane
73. n-Decanal
74. SMethyl-2-butanol
75. Propiophenone
76. Ethylacetate
77. n-Decane
78. Isopropylbenzene
79. IEthylpentane
80. Di-n-Butylamine
81. N-Dodecane
82. o-Dichlorobenzene
83. Allylacetate
84. S,SDiethylpentane
85. n-Butyl acetoacetate
86. Benzylamine
87. Indene
88. Methylnaphthalene
89. 'L-Methyl-Spentanone
90. Coumarin
91. Phenylacetic acid
92. Ethyl valerate
93. 5-Methyl-3-heptanone
94. n-Octane
95. Cumic alcohol
96. Methanol
97. 2,4-Dimethyl-Shexanone
98. Octylacetate
99. Cycloheptadiene
100. 2-Methyl-1-octene
101. Ethyl Lmethylvalerate
102. o-Nitrotoluene